Featured post

Four new elements added to periodic table

Four new elements added to periodic table The new elements – elements 113, 115, 117 and 118 – complete the periodic table’s seventh r...

Tuesday, 2 August 2016

Application of Graham Condenser

Reflux is the process of boiling reactants while continually cooling the vapor returning it back to the flask as a liquid. It is used to heat a mixture for extended periods and at certain temperatures...A condenser is attached to the boiling flask, and cooling water is circulated to condense escaping vapors.
If you are refluxing a mixture, as you might in organic synthesis to increase the speed of the reaction by doing it at a higher temperature (i.e., the boiling point of the solvent), then any of the condensers that worked well enough to avoid the loss of solvent and avoid "flooding" would work equally well. When you're refluxing, you want the "reflux ring", the place where the vapor is visibly condensing into a liquid, to be no more than 1/3 of the way up the reflux column.
You have two different basic types of condensers shown, Graham-type condensers (the first 3) and coil condensers (the last two). In the coil condensers (the left condenser in the picture below), the water flows through the coil and the vapor moves up in the larger, outside area of the condenser, condenses onto the cooled coils, then drips back into the pot. In a Graham-type condenser (the right condenser in the picture below), the water flows around a tube (whether straight or coiled) that contains the vapor/condensed liquid.(picture source) The Graham-type condensers clog (or flood) more easily since they have a more restricted path for the liquid to return to the pot.

enter image description here


Graham-type condensers: The Liebig condenser is simple, but has low cooling capacity and can be fairly easily clogged as the condensed liquid flows back into the flask and blocks the vapor that is trying to escape. The Allihn improves on this design by having a wider bore at the bottom and condensing the liquid on the "bubbles" where it can run down the sides and avoid blocking the vapor. (I've used this to good effect in refluxing many reactions.) The Graham condenser is the same basic design as the other two, but the condensation tube is coiled which provides more surface area for cooling...but also tends to send the condensed liquid right into the path of the vapor trying to move up. It is particularly prone to flooding.
Coil condensers, such as the Dimroth and Freidrichs, have high capacity for cooling with fewer problems from flooding since the vapor condenses on the coils and drips back from the little prominence at the bottom of the coils into the center of the pot. The vapor has an easy time getting past the drops falling into the pot. If you can afford it, this seems like a good choice for most applications. The Freidrichs condensers, which incorporate a cold-finger with the spiral, are higher capacity, quite bulky and heavy. I have seen them used with rotovaps where you are taking a lot of solvent off quickly, but not with an ordinary reflux apparatus. This would be over-kill for a simple reflux reaction situation.

Wednesday, 27 July 2016

Numerical

Guanosine has a maximum absorbance of 275 nm. ϵ275=8400M1cm1 and the path length is 1 cm. Using a spectrophotometer, you find the that A275=0.70.  What is the concentration of guanosine?
SOLUTION
To solve this problem, you must use Beer's Law.
                                               A=ϵlc
0.70 = (8400 M-1 cm-1)(1 cm)(c
Next, divide both side by [(8400 M-1 cm-1)(1 cm)]
                                                cc = 8.33x10-5 mol/L

Spectrophotometer

Spectrophotometry is a method to measure how much a chemical substance absorbs light by measuring the intensity of light as a beam of light passes through sample solution. The basic principle is that each compound absorbs or transmits light over a certain range of wavelength. This measurement can also be used to measure the amount of a known chemical substance. Spectrophotometry is one of the most useful methods of quantitative analysis in various fields such as chemistry, physics, biochemistry, material and chemical engineering and clinical applications.

Introduction

Every chemical compound absorbs, transmits, or reflects light (electromagnetic radiation) over a certain range of wavelength. Spectrophotometry is a measurement of how much a chemical substance absorbs or transmits. Spectrophotometry is widely used for quantitative analysis in various areas (e.g., chemistry, physics, biology, biochemistry, material and chemical engineering, clinical applications, industrial applications, etc). Any application that deals with chemical substances or materials can use this technique. In biochemistry, for example, it is used to determine enzyme-catalyzed reactions. In clinical applications, it is used to examine blood or tissues for clinical diagnosis. There are also several variations of the spectrophotometry such as atomic absorption spectrophotometry and atomic emission spectrophotometry.
A spectrophotometer is an instrument that measures the amount of photons (the intensity of light) absorbed after it passes through sample solution. With the spectrophotometer, the amount of a known chemical substance (concentrations) can also be determined by measuring the intensity of light detected. Depending on the range of wavelength of light source, it can be classified into two different types:
  • UV-visible spectrophotometer: uses light over the ultraviolet range (185 - 400 nm) and visible range (400 - 700 nm) of electromagnetic radiation spectrum.
  • IR spectrophotometer: uses light over the infrared range (700 - 15000 nm) of electromagnetic radiation spectrum.
In visible spectrophotometry, the absorption or the transmission of a certain substance can be determined by the observed color. For instance, a solution sample that absorbs light over all visible ranges (i.e., transmits none of visible wavelengths) appears black in theory. On the other hand, if all visible wavelengths are transmitted (i.e., absorbs nothing), the solution sample appears white. If a solution sample absorbs red light (~700 nm), it appears green because green is the complementary color of red. Visible spectrophotometers, in practice, use a prism to narrow down a certain range of wavelength (to filter out other wavelengths) so that the particular beam of light is passed through a solution sample. 

Devices and mechanism

Figure 1 illustrates the basic structure of spectrophotometers. It consists of a light source, a collimator, a monochromator, a wavelength selector, a cuvette for sample solution, a photoelectric detector, and a digital display or a meter. Detailed mechanism is described below. Figure 2 shows a sample spectrophotometer (Model: Spectronic 20D).
spectrophotometer_structure.png
Figure 1: Basic structure of spectrophotometers (illustrated by Heesung Shim)
A spectrophotometer, in general, consists of two devices; a spectrometer and a photometer. A spectrometer is a device that produces, typically disperses and measures light. A photometer indicates the photoelectric detector that measures the intensity of light.
  • Spectrometer: It produces a desired range of wavelength of light. First a collimator (lens) transmits a straight beam of light (photons) that passes through a monochromator (prism) to split it into several component wavelengths (spectrum). Then a wavelength selector (slit) transmits only the desired wavelengths, as shown in Figure 1.
  • Photometer: After the desired range of wavelength of light passes through the solution of a sample in cuvette, the photometer detects the amount of photons that is absorbed and then sends a signal to a galvanometer or a digital display, as illustrated in Figure 1.
Spc20Blnk.gif
Figure 2: A single wavelenth spectrophotometer
You need a spectrometer to produce a variety of wavelengths because different compounds absorb best at different wavelengths. For example, p-nitrophenol (acid form) has the maximum absorbance at approximately 320 nm and p-nitrophenolate (basic form) absorb best at 400nm, as shown in Figure 3.
wave.png
Figure 3: Absorbance of two different compounds
Looking at the graph that measures absorbance and wavelength, an isosbestic point can also be observed. An isosbestic point is the wavelength in which the absorbance of two or more species are the same. The appearance of an isosbestic point in a reaction demonstrates that an intermediate is NOT required to form a product from a reactant. Figure 4 shows an example of an isosbestic point.
 isopoint.png
Figure 4: An example of isosbestic point
Referring back to Figure 1 (and Figure 5), the amount of photons that goes through the cuvette and into the detector is dependent on the length of the cuvette and the concentration of the sample. Once you know the intensity of light after it passes through the cuvette, you can relate it to transmittance (T). Transmittance is the fraction of light that passes through the sample. This can be calculated using the equation:
Transmittance(T)=ItIo
Where It is the light intensity after the beam of light passes through the cuvette and Io is the light intensity before the beam of light passes through the cuvette. Transmittance is related to absorption by the expression:
Absorbance(A)=log(T)=log(ItIo)
Where absorbance stands for the amount of photons that is absorbed. With the amount of absorbance known from the above equation, you can determine the unknown concentration of the sample by using Beer-Lambert Law. Figure 5 illustrates transmittance of light through a sample. The length l is used for Beer-Lambert Law described below.
transmittance.png
Figure 5: Transmittance (illustrated by Heesung Shim)

Beer-Lambert Law

Beer-Lambert Law (also known as Beer's Law) states that there is a linear relationship between the absorbance and the concentration of a sample. For this reason, Beer's Law can only be applied when there is a linear relationship. Beer's Law is written as:
A=ϵlc
where
  • A is the measure of absorbance (no units),
  • ϵ is the molar extinction coefficient or molar absorptivity (or absorption coefficient),
  • l is the path length, and
  • c is the concentration.
 The molar extinction coefficient is given as a constant and varies for each molecule. Since absorbance does not carry any units, the units for ϵ must cancel out the units of length and concentration. As a result, ϵ has the units: L·mol-1·cm-1. The path length is measured in centimeters. Because a standard spectrometer uses a cuvette that is 1 cm in width, l is always assumed to equal 1 cm. Since absorption, ϵ, and path length are known, we can calculate the concentration c of the sample.